Beurer BM 26 Bedienungsanleitung Seite 78

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 80
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 77
78
Electromagnetic Compatibility Information
Table 1
Guidance and manufacture’s declaration – electromagnetic emissions
The BM 26 is intended for use in the electromagnetic environment specifi ed below.
The customer or the user of the BM 26 should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment – guidance
RF emissions CISPR 11 Group 1 The BM 26 uses RF energy only for its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions CISPR 11 Class B The BM 26 is suitable for use in all establishments other than domestic and those
directly connected to the public low-voltage power supply network that supplies buil-
dings used for domestic purposes.
Harmonic emissions IEC 61000-3-2 Not applicable
Voltage fl uctuations/fl icker emissions
IEC 61000-3-3
Not applicable
Table 2
Guidance and manufacturer’s declaration – electromagnetic immunity
The BM 26 is intended for use in the electromagnetic environment specifi ed below. The customer or the user of the BM 26 should assure that it
is used in such an environment.
IMMUNITY test IEC 60601 test level Compliance level Electromagnetic environment – guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or ceramic tile. If fl oors are covered
with synthetic material, the relative humidity should be at least 30 %.
Power frequency (50/60 Hz)
magnetic fi eld IEC 61000-4-8
3 A/m 3 A/m
Power frequency magnetic fi elds should be at levels characteristic of
a typical location in a typical commercial or hospital environment.
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.
Seitenansicht 77