Beurer GL32 mg/dL
18
11
Carefully remove the needle lancet from the lancing
device and discard the lancet in a suitable sharp-proof
container.
Carefully dispose of all blood samples and materials
with which you have come into contact. This prevents
injury and prevents infection spreading to other people.
12
Twist the (grey) cap back in place.
5.2 Measuring the blood sugar value
1
Hold the measuring device so that the display is facing you.
2
Insert a test strip, contacts first, into the device. Ensure that the front side is
facing you. You can handle any part of the test strip with clean, dry hands.
3
The device switches on automatically and shows the start display, shortly
followed by the checking symbol “
“ . As soon as the flashing blood droplet
symbol 6 is displayed, the device is ready to perform measurements.
Hold the blood collection slit (at the point of the test
strip) against the blood droplet. Do not press the
puncture site (fingertip or other body part) against the
test strip. The blood must not be smeared. The blood
is sucked into the slit.
4
WARNING
Hold the blood collection slit of the test strip against the blood
droplet until the slit is completely filled and you hear a beep.
Withdrawing the test strip from the blood droplet before the beep
may lead to incorrect measurements.
5
Once the slit is filled with blood, the device performs the blood sugar meas-
urement. The measuring device counts down for 6 seconds, and the result of
the measurement is then shown in the display. Read the measured value. For
an explanation of the measured value and possible actions, see the next
chapter "5.3 Evaluating the measured value", page 19.
If an error message is displayed, refer to chapter “8 Troubleshooting" on page 29.